Skip to main content
VICChemistrySyllabus dot point

How are substances in water measured and analysed?

the principles and use of colorimetry and UV-visible spectroscopy (including the Beer-Lambert relationship) and atomic absorption spectroscopy (AAS), and the use of calibration curves to determine the concentration of an analyte in water

A focused VCE Chemistry Unit 2 answer on instrumental analysis. Covers the principles of colorimetry and UV-visible spectroscopy with the Beer-Lambert relationship, the use of calibration curves, and atomic absorption spectroscopy (AAS) for trace-metal analysis, with a comparison of techniques.

Generated by Claude Opus 4.810 min answer

Reviewed by: AI editorial process; not yet individually human-reviewed

Have a quick question? Jump to the Q&A page

Jump to a section
  1. What this dot point is asking
  2. The answer
  3. Examples in context
  4. Try this

What this dot point is asking

VCAA wants you to describe the principles of colorimetry, UV-visible spectroscopy and atomic absorption spectroscopy, to apply the Beer-Lambert relationship (A = ecl), to use a calibration curve to determine the concentration of an analyte, and to compare the techniques for suitability.

The answer

Why use instrumental analysis

Volumetric methods are accurate but limited to analytes that react in a known stoichiometry and at concentrations high enough to titrate. Instrumental methods extend the toolkit to:

  • Coloured solutions (where the colour itself is the analyte).
  • Trace metals at ppm or ppb levels (well below titration's reliable range).
  • Mixtures where one component absorbs at a wavelength others do not.

Common principle: pass light through the sample, measure how much is absorbed at a chosen wavelength, and convert absorbance to concentration using a calibration curve.

Colorimetry and UV-visible spectroscopy

Coloured solutions absorb visible light because their electrons are promoted between energy levels separated by a visible-light photon. The colour of a solution is the complement of the colour it absorbs (a blue solution absorbs orange/red; a red solution absorbs green).

A colorimeter passes a narrow band of visible light (often filtered) through the sample. A UV-visible spectrophotometer scans across a range of wavelengths (often 200 to 800 nm) and produces an absorbance spectrum.

The Beer-Lambert relationship quantifies the absorbance:

A = e x c x l

  • A is absorbance (dimensionless; log10 of I0/I).
  • e is the molar absorptivity (or extinction coefficient), characteristic of the analyte at that wavelength, in L mol^-1 cm^-1.
  • c is the concentration of the absorbing species, in mol L^-1.
  • l is the path length of the cuvette, in cm (usually 1.00 cm).

For a given analyte and cuvette, A is proportional to c at low concentration: A = constant x c. This is exactly what a calibration curve relies on.

Calibration curves

The standard workflow:

  1. Prepare a series of standards of known concentration spanning the expected range of the unknown.
  2. Measure absorbance for each at a chosen wavelength (usually the wavelength of maximum absorbance, lambda_max).
  3. Plot absorbance (y) against concentration (x). At low concentration the plot is linear and passes through (or near) the origin.
  4. Measure the absorbance of the unknown.
  5. Read the concentration of the unknown directly from the line (do not extrapolate beyond the calibrated range).

Why use the wavelength of maximum absorbance? Two reasons:

  • Sensitivity: the calibration curve has its steepest slope, so a small change in c produces the largest change in A.
  • Robustness: small wavelength drifts in the instrument do not change A much (the peak is locally flat).

Atomic absorption spectroscopy (AAS)

AAS is used for metal ions in water at trace concentrations (ppm or ppb). The principle: free, gaseous, ground-state metal atoms absorb light at the same wavelengths their atoms emit. Each element has a unique line spectrum, so a specific lamp gives a specific element's lines.

The instrument:

  1. Hollow-cathode lamp: a tube containing the same element being analysed. A current through it excites atoms of that element, which emit only the wavelengths characteristic of that element. (A Pb lamp emits Pb wavelengths only.)
  2. Atomiser (a flame, typically air/acetylene): aspirates the sample, evaporates the solvent, breaks compounds into atoms, and reduces ions to ground-state atoms.
  3. Monochromator: selects one specific wavelength (usually the strongest line).
  4. Detector: measures the intensity of light that passes through the flame.

The absorbance is proportional to the number density of ground-state atoms of that element in the flame, which is proportional to the concentration in the original sample.

Calibration is identical to UV-visible: a series of standards of known concentration of the same element, measured under the same conditions, gives a linear calibration curve.

Strengths of AAS:

  • Element-specific (the lamp ensures only the target element is measured).
  • Sensitive to ppm/ppb.
  • Many elements can be analysed (Cu, Fe, Pb, Cd, Zn, Hg, etc., one at a time per lamp).

Limitations:

  • One element at a time (multiple elements require multiple lamps).
  • Not suitable for non-metals or for total organic content.
  • Sample matrix (other species in the water) can interfere; matrix-matched standards help.

Comparison of techniques

Technique Best for Detection limit Notes
Colorimetry Coloured species, classroom labs mmol L^-1 to mol L^-1 Cheap, robust, single filter
UV-visible Coloured species, organic chromophores, more sensitive than colorimetry umol L^-1 to mmol L^-1 Variable wavelength, full spectrum
AAS Trace metal ions in water ppm to ppb (umol L^-1 to nmol L^-1) Element-specific via lamp choice
Titration (for comparison) Bulk concentrations, matched to a reaction mmol L^-1 to mol L^-1 Requires a known stoichiometry

Examples in context

Example 1. Lead in Mount Isa kindergarten dust by AAS. Public health chemists at the National Measurement Institute test settled dust from kindergartens within 2km2 \, \text{km} of the Mount Isa lead smelter using flame AAS. A wipe sample is digested in concentrated HNO3\text{HNO}_3 and made up to 25.0mL25.0 \, \text{mL}. A lead hollow-cathode lamp emits at 217.0nm217.0 \, \text{nm}; the sample absorbance is 0.4120.412 against calibration standards of 1,2,51, 2, 5 and 10ppm10 \, \text{ppm} giving slope 0.0825ppm10.0825 \, \text{ppm}^{-1}. Concentration in solution =0.412/0.0825=5.00ppm= 0.412 / 0.0825 = 5.00 \, \text{ppm}. Mass of lead =5.00mg/L×0.0250L=0.125mg= 5.00 \, \text{mg/L} \times 0.0250 \, \text{L} = 0.125 \, \text{mg}. Against the 25μg25 \, \mu\text{g} per 100cm2100 \, \text{cm}^2 NHMRC reference, the wipe (from 0.5m20.5 \, \text{m}^2) is six times the threshold.

Example 2. Iron in Yarra Valley winery storage tanks by UV-vis. Winemakers monitor dissolved iron because levels above 5ppm5 \, \text{ppm} catalyse oxidation that browns chardonnay. A wine sample is buffered to pH 3.53.5 and reacted with 1,101,10-phenanthroline to form a red Fe(phen)32+\text{Fe(phen)}_3^{2+} complex with maximum absorbance at 510nm510 \, \text{nm}. With path length b=1.00cmb = 1.00 \, \text{cm} and molar absorptivity ϵ=11,100Lmol1cm1\epsilon = 11{,}100 \, \text{L} \cdot \text{mol}^{-1} \cdot \text{cm}^{-1}, an absorbance of 0.4980.498 gives c=A/(ϵb)=0.498/11,100=4.49×105mol/Lc = A / (\epsilon b) = 0.498 / 11{,}100 = 4.49 \times 10^{-5} \, \text{mol/L}. Multiplying by M(Fe)=55.8M(\text{Fe}) = 55.8 converts to 2.50mg/L=2.50ppm2.50 \, \text{mg/L} = 2.50 \, \text{ppm}, well within specification.

Try this

Q1. State the wavelength chosen for analysis in UV-vis spectroscopy and explain why this choice gives the most accurate result. [2 marks]

  • Cue. The wavelength of maximum absorbance (λmax\lambda_{max}). Small wavelength errors give smallest change in absorbance, so the calibration is most sensitive there.

Q2. Four copper standards of 1.00,2.00,4.001.00, 2.00, 4.00 and 6.00ppm6.00 \, \text{ppm} give absorbances of 0.085,0.170,0.3400.085, 0.170, 0.340 and 0.5100.510. A sample reads 0.2550.255. Calculate the copper concentration in the sample. [3 marks]

  • Cue. Slope of calibration 0.0850ppm1\approx 0.0850 \, \text{ppm}^{-1}; c=0.255/0.0850=3.00ppmc = 0.255 / 0.0850 = 3.00 \, \text{ppm} of Cu.

Q3. A water sample is analysed by AAS for zinc. The dilution factor is ×10\times 10; the cuvette absorbance is 0.2200.220 on a calibration with slope 0.044ppm10.044 \, \text{ppm}^{-1}. (a) Calculate cuvette [Zn][\text{Zn}]. (b) Calculate the original [Zn][\text{Zn}]. (c) Compare AAS and UV-vis for this analysis. [2+2+2 marks]

  • Cue. (a) 5.00ppm5.00 \, \text{ppm}. (b) 50.0ppm50.0 \, \text{ppm}. (c) AAS uses element-specific hollow-cathode lamp on atomised sample; very low detection limit for metals; UV-vis needs coloured complex. Zinc is colourless, so AAS preferred.

Exam-style practice questions

Practice questions written in the style of VCAA exam questions on this dot point, with worked answer explainers. The year tag is the paper they imitate, not the source.

2024 VCE4 marksA series of copper(II) sulfate standards gives the following absorbances at 600 nm: 0.100 mol L^-1 -> 0.250; 0.200 mol L^-1 -> 0.500; 0.300 mol L^-1 -> 0.750; 0.400 mol L^-1 -> 1.000. An unknown solution has an absorbance of 0.620. (a) Sketch the calibration curve and (b) determine the concentration of the unknown. (c) Why is 600 nm the chosen wavelength?
Show worked answer →

A 4-mark answer needs the linear curve, the unknown concentration by interpolation, and the wavelength justification.

(a) Plot absorbance (y) against concentration (x). The four points fall on a straight line through the origin with gradient 2.50 L mol^-1 (each 0.100 mol L^-1 rise gives an absorbance rise of 0.250). The line obeys the Beer-Lambert relationship A = ecl with the slope equal to el.

(b) From the line: c = A / gradient = 0.620 / 2.50 = 0.248 mol L^-1.

Alternatively interpolate between (0.200, 0.500) and (0.300, 0.750): linear interpolation gives 0.200 + 0.100 x (0.620 - 0.500)/(0.750 - 0.500) = 0.248 mol L^-1.

(c) 600 nm is the wavelength of maximum absorbance for the blue Cu(H2O)6^2+ ion (which appears blue because it absorbs in the orange/red part of the spectrum). Using the wavelength of maximum absorbance maximises sensitivity (largest A for a given c) and means small wavelength errors do not change A much (the absorbance peak is flat at its maximum).

2025 VCE4 marksA water sample is suspected to contain trace amounts of lead. (a) Why is AAS more appropriate than UV-visible spectroscopy for this determination? (b) Outline how AAS is calibrated and how the lead concentration is calculated.
Show worked answer →

A 4-mark answer needs the suitability argument and the calibration procedure.

(a) Why AAS, not UV-visible: lead at trace level (ppm or ppb) does not produce a strong, characteristic colour in solution; UV-visible would not have the sensitivity. AAS is element-specific (a hollow-cathode lamp emits only Pb wavelengths), highly sensitive (ppb-level detection limit), and not interfered with by colour from other species.

(b) AAS procedure:

  1. Atomise the sample in a flame. Ground-state atoms of every element are produced.
  2. Pass light from a lead hollow-cathode lamp (which emits only Pb-characteristic wavelengths) through the flame. Ground-state Pb atoms absorb the light at their specific wavelengths.
  3. Measure the absorbance.
  4. Prepare a series of standard solutions of known Pb concentration. Measure absorbance for each. Plot absorbance vs concentration to make the calibration curve (should be linear at low concentration).
  5. Measure absorbance of the water sample. Read its lead concentration from the calibration curve.

Markers also accept comments on matrix matching (using a similar background matrix in the standards to that of the sample) and on running blanks to subtract any baseline absorbance.

Related dot points